Backing pump capacity |
15 m³/h |
Detectable gases |
4He, 3He, H2
|
Flange (in) |
DN 25 ISO-KF |
I/O interfaces |
6 Digital inputs (Allocated functions configurable); 3 Analog outputs (Configurable : Helium signal log, Mantissa, Exponent, Inlet pressure); 5 Relays outputs (Allocated functions configurable); 4 Transitor/open collector outputs (Allocated functions configurable) |
Interface |
RS-232, USB |
Max. inlet test pressure |
25 hPa |
Minimum detectable leak rate for helium (sniffing leak detection) |
5 · 10-10 Pa m3/s |
Minimum detectable leak rate for helium (vacuum leak detection) |
5 · 10-13 Pa m3/s |
Noise level |
54 dB(A) |
Operating temperature (hard vacuum test) |
0-45 °C |
Operating temperature (sniffing test) |
0-35 °C |
Power consumption max. |
850 W |
Protection category |
IP 20 |
Pumping speed for He |
2.5 l/s |
Start-up time (20°C) without calibration |
~ 3 min |
Supply |
110-130 V, 50/60 Hz |
Test method |
Vacuum and sniffing leak detection |
Warranty: Year(s) |
2 |
Weight |
56 kg | 123.46 lb |