| Backing pump capacity |
15 m³/h |
| Detectable gases |
4He, 3He, H2
|
| Flange (in) |
DN 25 ISO-KF |
| I/O interfaces |
6 Digital inputs (Allocated functions configurable); 3 Analog outputs (Configurable : Helium signal log, Mantissa, Exponent, Inlet pressure); 5 Relays outputs (Allocated functions configurable); 4 Transitor/open collector outputs (Allocated functions configurable) |
| Interface |
RS-232, USB |
| Max. inlet test pressure |
25 hPa |
| Minimum detectable leak rate for helium (sniffing leak detection) |
5 · 10-10 Pa m3/s |
| Minimum detectable leak rate for helium (vacuum leak detection) |
5 · 10-13 Pa m3/s |
| Noise level |
54 dB(A) |
| Operating temperature (hard vacuum test) |
0-45 °C |
| Operating temperature (sniffing test) |
0-35 °C |
| Power consumption max. |
850 W |
| Protection category |
IP 20 |
| Pumping speed for He |
2.5 l/s |
| Start-up time (20°C) without calibration |
~ 3 min |
| Supply |
110-130 V, 50/60 Hz |
| Test method |
Vacuum and sniffing leak detection |
| Warranty: Year(s) |
2 |
| Weight |
56 kg | 123.46 lb |